3M™ Comply™ Sterilization Bowie-Dick Plus Test Pack
724429CS

$0.00

  • Designed to monitor the performance of vacuum-assisted 270° – 273° F steam sterilizers each day
  • An early warning sheet that detects potential issues before they appear on the test sheet
  • Early warning feature of the Comply Bowie-Dick Plus Test Pack permits scheduled sterilizer maintenance and helps to prevent unexpected failures
  • Easy-to-read test sheet clearly identifies failures and can be filed for record keeping
  • Lead free ink chemistry

340 in stock

Description

A Bowie-Dick test should be run daily on steam sterilizers. It should also be used to check sterilizer performance after a major repair. The 3M Comply Bowie-Dick Plus Test Pack 00135LF is designed for testing air removal efficiency of 270˚ – 273˚ F dynamic-air-removal steam sterilizers. It features an early warning sheet that is designed for early detection of air removal failures before they appear on the primary test sheet. Test packs will detect residual air in the sterilizer chamber due to air leaks, air removal failures and inadequate vacuum.